Quality Control of a Laser Additive Manufactured Medical Implant by X-Ray Tomography

dc.contributor.authordu Plessis, Anton
dc.contributor.authorle Roux, Stephan Gerhard
dc.contributor.authorBooysen, Gerrie
dc.contributor.authorEls, Johan
dc.date.accessioned2017-11-16T11:45:22Z
dc.date.available2017-11-16T11:45:22Z
dc.date.issued2016
dc.descriptionPublished Articleen_US
dc.description.abstractQuality control of laser additive manufactured medical implants is of interest, especially if nondestructive quality control can be performed on parts before implantation. X-ray micro-computed tomography (microCT or CT) can be used for defect/porosity analysis as well as for comparing the part surface with its computer-aided design (CAD) file. In both cases, the limited use of CT is partly due to the variation in scan types and the quality of scans that can occur. We present a simple method demonstrating the use of a light metal casting as a reference porosity sample, to confirm good CT image quality and to quantify minimum detectable pore size for the selected CT scan settings. This makes a good comparison for additive manufactured parts, since castings generally contain more porosity. A full part-to-CAD comparison shows how the part is compared with its CAD file, as a second-quality control. The accuracy of the CAD variance is given by the minimum detectable pore size. Finally, the part is sectioned and scanned at two higher resolution settings showing small porosity (10–50 lm diameter) present but well distributed, as expected.en_US
dc.format.extent766 444 bytes, 1 file
dc.format.mimetypeApplication/PDF
dc.identifier.urihttp://hdl.handle.net/11462/1266
dc.language.isoen_USen_US
dc.publisherStellenbosch University: 3D Printing and Additive Manufacturingen_US
dc.relation.ispartofseriesVolume: 3 Issue 3;
dc.subjectX-ray tomographyen_US
dc.subjectadditive manufacturingen_US
dc.subjectporosityen_US
dc.subjectdefect detectionen_US
dc.subjectmedical implanten_US
dc.subjectquality controlen_US
dc.titleQuality Control of a Laser Additive Manufactured Medical Implant by X-Ray Tomographyen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
G Booysen 3.pdf
Size:
748.48 KB
Format:
Adobe Portable Document Format
Description:
Published Article

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: